Besides adequate steam penetration, which can be measured with the 4D Sensor, it is also of great importance that the quality of the steam and conditions in the chamber are measured in every sterilization process. The NCG sensor measures in real-time the exact value of non-condensable gases in the chamber during every sterilization process, providing operators with the most reliable and efficient steam sterilization processes possible.
With the measured percentage of NCGs and chamber temperature the sterilization conditions, as specified in EN 285:2015, can be verified. With this evidence, the sterilized load can be released for use; parametric release based upon directly measured sterilization parameters. The NCG sensor is mounted directly on the chamber and is maintenance-free. It further reduces the need for separate test cycles, resulting in increased sustainability, less consumption, reduced waste, and the need for staff to perform separate tests.